Table 16: Thermocouple Read and Source (errors in °C)
TC Type Range (°C) Accuracy
J -210.0 - 0.0 0.4
0.0 - 800.0 0.2
800.0 - 1200.0 0.3
K -200.0 - 0.0 0.6
0.0 - 1000.0 0.3
1000.0 - 1372.0 0.5
T -250.0 - 0.0 0.6
0.0 - 400.0 0.2
E -250.0 - -100.0 0.6
-100.0 - 1000.0 0.2
R 0.0 - 1767.0 1.2
S 0.0 - 1767.0 1.2
B 600.0 - 800.0 1.2
800.0 - 1000.0 1.3
1000.0 - 1820.0 1.5
C 0.0 - 1000.0 0.6
1000.0 - 2316.0 2.3
XK -200.0 - 800.0 0.2
BP 0.0 - 800.0 0.9
800.0 - 2500.0 2.3
L -200.0 - 0.0 0.25
0.0 - 900.0 0.2
U -200.0 - 0.0 0.5
0.0 - 600.0 0.25
N -200.0 - 0.0 0.8
0.0 - 1300.0 0.4
All TC errors include CJC errors
CJC error outside of 23 ± 5°C is 0.05°C/°C
(In °C add .2 for cold junction compensation error.)
Table 17: RTD Read and Source
RTD Type Range (°C) Uncertainty
Miniumum Maximum 1 year (°C)
PT385, 10 ohm -200 -80 0.78
-80 0 0.82
0 100 0.84
100 300 0.95
300 400 0.99
400 630 1.09
630 800 1.19
99