Operating Instructions
3-6
If no key is pressed for 10 seconds, the instrument will save the current configuration
and exit the Setup Menu.
View Modes and Statistics
The instrument features multiple view modes including a variety of real-time statistics, all
available at the touch of a button. Figure 6 below describes each view mode.
Set 0% Level
Range (rAnGE)
Range Hi mV (default)
[-15mV to +85mV]
Range Lo mV
7
(mV flashing)
[-15mV to +35mV]
If on Set Mode
Manual
blinking
“__ __ __ __”
Fast Ramp
Slow Ramp
Step
Transfer
7
Low range is for calibration verification only.
Figure 5
VIEW MODE
D
ISPLAY
INDICATOR
DESCRIPTION
Minimum MIN
Minimum temperature recorded during current
session
Maximum MAX
Maximum temperature recorded during current
session
Average AVG
Average of all temperatures recorded during
current session
Range RNG Maximum minus Minimum
Standard Deviation STDEV
Standard deviation of all temperatures recorded
during the current session
1
.
1
Standard Deviation is calculated using the population formula: =
(−µ)
2
Figure 6: View Modes and Statistics