EasyManua.ls Logo

Omega DPG4000 - Page 14

Omega DPG4000
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Use the Pressure Standard to output 0.00 psi. After the output has
stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
Note: Only some ranges use vacuum calibration. If your gauge is not
one, than this step will be automatically skipped and calibration will be
complete.
Use the Pressure Standard to output -12.00 psi. After the output has
stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show ———-. When the readings are
complete the unit will reset and power up in normal mode.
PSI
-12.00
CAL MODE
PSI
0.00
CAL MODE

Related product manuals