Use the Pressure Standard to output 312.50 psi (62.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 250.00 psi (50%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 187.50 psi (37.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 125.00 psi (25%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 62.50 psi (12.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
11
PSI
250.00
CAL MODE
PSI
187.50
CAL MODE
PSI
125.00
CAL MODE
PSI
62.50
CAL MODE