Bead Ruptor 24 Elite
BEAD RUPTOR 24 ELITE
Bead Mill Homogenizer
en User Manual
fr Manuel d’utilisation
it Manuale d’uso
es Manual del usuario
de Benutzerhandbuch

Questions and Answers

  • J
    James GouldAug 7, 2025
    What to do if the fan is not working on my Omni Bead Ruptor 24 Elite Laboratory Equipment?
    • M
      Michelle KimAug 7, 2025
      If the fan in your Omni Laboratory Equipment is not working, start by checking the main voltage and ensuring it matches the unit's requirements. Make sure the unit is properly plugged in and replace any blown fuses. If the fan still doesn't work, it might be faulty, so turn off the unit and contact technical support.
  • A
    Aaron KruegerAug 11, 2025
    What to do if there is no display on the screen of my Omni Bead Ruptor 24 Elite Laboratory Equipment?
    • K
      Kathleen OsbornAug 11, 2025
      If there is no display on the screen of your Omni Laboratory Equipment, first check the main voltage and ensure the unit's voltage matches the power supply. Verify the unit is properly plugged in and replace any blown fuses. If the problem continues, it may be due to a faulty display screen, in which case you should turn off the unit and contact technical support.
  • K
    Kim BartonAug 15, 2025
    What to do if there is tube leakage in my Omni Bead Ruptor 24 Elite?
    • J
      Jared LewisAug 16, 2025
      If you're experiencing tube leakage with your Omni Laboratory Equipment, it might be due to excessive speed, or high temperature or pressure inside the tubes. Try reducing the processing time or speed to resolve this issue.
  • D
    Daniel JimenezAug 20, 2025
    How to fix incomplete characters on the display screen of my Omni Bead Ruptor 24 Elite Laboratory Equipment?
    • W
      william74Aug 20, 2025
      If the display screen on your Omni Laboratory Equipment shows incomplete or unusual characters, the CPU may need to be reset. Turn off the unit, wait 10 seconds, and then turn it back on.
  • I
    ilawrenceAug 25, 2025
    Why does my Omni Bead Ruptor 24 Elite Laboratory Equipment power on, but the motor does not turn?
    • R
      Robert WalkerAug 25, 2025
      If your Omni Laboratory Equipment powers on but the motor doesn't turn, the incorrect voltage may be selected. Power off the unit and select the appropriate voltage. If the problem persists, contact Omni International for assistance.
  • D
    Dustin SmithAug 28, 2025
    How to disengage a difficult fingerplate lock on my Omni Bead Ruptor 24 Elite?
    • D
      Denise Byrd MDAug 28, 2025
      If the fingerplate lock on your Omni Laboratory Equipment is difficult to disengage, turn the knob slightly counter-clockwise to release tension.
  • N
    Nicholas GreenSep 2, 2025
    What does it mean if my Omni Bead Ruptor 24 Elite powers on with a loud bang?
    • S
      Sharon ParkSep 2, 2025
      If your Omni Laboratory Equipment powers on with a loud “BANG”, this indicates the motor control device is damaged due to incorrect voltage selection. Return the unit to Omni International for replacement of the motor control device.

Omni Bead Ruptor 24 Elite Specifications

General IconGeneral
ModelBead Ruptor 24 Elite
Bead Mill TypeHigh-energy bead mill
Sample CapacityUp to 24 samples
Number of Cycles1 to 10
Weight44 lbs (20 kg)
Power Requirements100-240 VAC, 50/60 Hz
Capacity24 tubes
Tube Sizes2.0 mL
Time Range1 second to 9 minutes 59 seconds
Cooling OptionOptional
Bead Tube Volume0.5 mL