EasyManuals Logo

Omron BP761CAN User Manual

Omron BP761CAN
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
32
9. Guidance and Manufacturer’s Declaration
Guidance and manufacturers declaration - electromagnetic immunity
OMRON BP761CAN including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of
this OMRON BP761CAN including AC-adapter should assure that it is used in such environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a typical
commercial and/or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a typical
commercial and/or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply inputlines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0.5 cycle
<5 % UT (>95 % dip in UT)
for 0.5 cycle
Mains power quality should be that of a typical
commercial and/or hospital environment. If the
user of the OMRON BP761CAN including AC-
adapter requires continued operation during
power mains interruption, it is recommended
that the OMRON BP761CAN including AC-
adapter be powered from an uninterruptible
power supply.
40 % UT (60 % dip in UT)
for 5 cycles
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (95 % dip in UT)
for 5 sec.
<5 % UT (95 % dip in UT)
for 5 sec.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note: UT is the A.C. mains voltage prior to application of the test level.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron BP761CAN and is the answer not in the manual?

Omron BP761CAN Specifications

General IconGeneral
BrandOmron
ModelBP761CAN
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals