26
Sample 2 99.6 mg/dL(5.5mmol/L) 3.0%
Sample 3 152.4 mg/dL(8.5mmol/L) 2.8%
Sample 4 247.8 mg/dL(13.8mmol/L) 2.9%
Sample 5 400.0 mg/dL(22.2mmol/L) 3.1%
Between Day Precision
Control solution 1 56.2 mg/dL(3.1mmol/L)
SD=4.2 mg/dL (0.2
mmol/L)
Control solution 2 104.1 mg/dL(5.8mmol/L) 3.2%
Control solution 3 306.1 mg/dL(17.0mmol/L) 2.9%
Chapter 9 Specication and Maintenance
Measuring range: 20~600mg/dL or 1.1~33.3mmol/L
Measuring unit: mg/dL or mmol/L
Measuring time: Approximately 5 seconds
Blood Sample size: Approximately 1μL
Operating temperature: (10
~40
)/(50 ºF ~104ºF)
Store the test strips at ambient temperature between
4~30
(39ºF ~86ºF). Do not refrigerate.
Relative operating humidity: Below 90% RH
Relative storage humidity: Below 75% RH
Memory capacity: Up to 512 test results (together with
date, time and unit) include average results for the last 7,
14 and 28 days
Dimension: 9.0 x 6.0 x 2.5 cm (L×W×H)
Weight: Approximately 90g with batteries
Power supply: Two 1.5V AAA batteries
Recommend to store in room temperature environment.
Please do not store in high temperature or high humidity.
Please prevent the HEA-230 system against from impact.
Please do not disassemble tthe HEA-230 system.
warning: The measurement unit either mg/dL or mmol/L is xed in the
meter according to the country demand. You are not able to change the
unit in the meter.
Caution: Any further service and product information, please contact
OMRON distributors.