10
IRREGULAR HEARTBEAT SYMBOL( )
When the monitor detects an irregular rhythm two or more times during
the measurement, the Irregular Heartbeat Symbol ( ) will appear on
the display with the measurement values.
An irregular heartbeat rhythm is defined as
a rhythm that is more than 25 % slower
or 25 % faster from the average rhythm
detected while the monitor is measuring
the systolic blood pressure and the
diastolic blood pressure.
If the Irregular Heartbeat Symbol ( )
displays with your measurement results,
we recommend you consult your
physician. Follow the directions of your
doctor.
AVERAGE VALUE SYMBOL ( )
The Average Value Symbol is displayed if the measurement values
displayed are an average of up to three measurements taken within 10
minutes of the last reading.
MOVEMENT ERROR SYMBOL ( )
The Movement Error Symbol is displayed if you move your body during
the measurement. Please remove the arm cuff, and wait 2-3 minutes.
Take another measurement.
DISPLAY SYMBOLS
Normal Heartbeat
Irregular Heartbeat
Pulse
Blood
pressure
Pulse
Blood
pressure
Short Long