RTD Transducer K3FP-RS-UI 29
Precautions
Refer to pages 38 and 39 for common precautions.
Precautions for Correct Use
Diagnostics Indicator
The diagnostics indicator (LED) inside the translucent cover shows
the error status, as outlined in the following table.
Connections
2-conductor Connection
For short distances (less than 10 m)
Cable resistances R
L1
and R
L2
are incorporated in the
measurement result directly and falsify the result accordingly.
3-conductor Connection
For long distances between PT100 platinum resistance
thermometers and K3FP-RS-UI RTD Transducers.
The value of all cable resistances must be exactly the same to
balance out the PT100 platinum resistance thermometer cable
resistances (R
L1
= R
L2
= R
L3
).
4-conductor Connections
For long distances between the PT100 platinum resistance
thermometer and the K3FP-RS-UI and different cable resistances
(R
L1
R
L2
R
L3
R
L4
).
Indicator status Error details
Flashing Measuring range span less than 50 K
Lit Disconnection on the platinum resistance
thermometer side
Short circuit on the platinum resistance
thermometer side
Above measurement range
Below measurement range
1
2
Inputs
3
4
R
L2
RL1
1
2
3
4
Inputs
RL1
RL3
RL2
1
2
3
4
Inputs
RL1
RL4
RL3
RL2