ERROR MESSAGES AND DETAILS ABOUT YOUR SYSTEM
No result in memory, such as the rst time
use of the meter.
or,
Your meter was unable to recall this result.
You can still perform a blood glucose test
and get an accurate test result. Contact your
local LifeScan Representative to report this
occurrence if this is not your rst time use
of the meter.
Error message indicates there is a problem
with the meter.
Do Not use the meter. Contact your local
LifeScan Representative.
Error message could be caused either by a
used test strip or a problem with the meter.
Repeat the test with a new test strip; see
Section3, Testing your blood glucose. If this
message continues to appear, contact your
local LifeScan Representative.
Error message indicates that the blood or
control solution sample was applied before
the meter was ready.
Repeat the test with a new test strip. Apply
a blood or control solution sample only after
the ashing blood drop icon
R
appears
on the display. If this message continues
to appear, contact your local LifeScan
Representative.