ERROR MESSAGES AND DETAILS ABOUT YOUR SYSTEM
The meter has detected a problem with the test
strip. Possible causes are test strip damage or
an incompletely lled conrmation window.
Repeat the test with a new test strip. Refer to
blood application (see Section 3, Testing your
blood glucose) or control solution testing (see
Section 5, Control solution testing).
Meter battery is low but there is enough power
to perform a test.
When the ashing battery icon rst appears,
there is enough power for a minimum of
100 more tests. Test results will still be
accurate, but replace the battery as soon
as possible.
The
icon ashes on the display by itself when
there is not enough battery power to perform a
test or view previous results.
Replace the meter battery.
Message What it means What to do
One of the following may apply:
You may have high glucose and have
tested in an environment near the
low end of the system’s operating
temperature range (6–44°C).
or,
There may be a problem with the test
strip. For example, it may have been
damaged or moved during testing.
or,
The sample was improperly applied.
or,
There may be a problem with the
meter.
If you tested in a cool environment, repeat the
test in a warmer environment with a new test
strip; see Section 3, Testing your blood glucose.
If this error message appears again, contact your
local LifeScan Representative.
If you tested in a normal or warm environment,
repeat the test with a new test strip; see
Section 3, Testing your blood glucose. If this
error message appears again, contact your local
LifeScan Representative.
If you applied the sample incorrectly, review
blood application (see Section 3, Testing your
blood glucose) or control solution testing (see
Section 5, Control solution testing) and repeat
the test with a new test strip. If the error
message appears again, contact your local
LifeScan Representative.
If this error message appears again, contact your
local LifeScan Representative.