40
You may have high
glucose and have tested
in an environment near
the low end of the
system’s operating
temperature range
(6–44
°
C).
or,
see top of next page
If you tested in a cool
environment, repeat the test in a
warmer environment with a new
test strip; see pages 22–23. If
this error message appears
again, contact Customer Care.
Message What it means What to do
One of the following may apply:
Understanding
error messages
Error message indicates
that the blood or control
solution sample was
applied before the meter
was ready.
Repeat the test with a new test
strip. Apply a blood or control
solution sample only after the
flashing blood drop icon
R
appears on the display. If this
message continues to appear,
contact Customer Care.
DCO_06397001C_OTUE_OB_GB.qxd 4/13/07 3:18 PM Page 40