38
4.1 Process interface
The PI is equipped with a process interface (cable with integrated electronics and terminal block), which
can be programmed via the software as an Analog Input (AI) and Digital Input (DI) in order to control the
camera or as an Analog Output (AO) in order to control the process. The signal level is always 0-10 V.
The process interface can be activated choosing the following options:
Analog Input (AI):
Emissivity, ambient temperature, reference temperature, flag control, triggered recording, triggered
snapshots, triggered line-scanner, uncommitted value
Analog Output (AO):
Main area temperature, internal temperature, flag status, alarm
Digital Input (DI):
Flag control, triggered recording, triggered snapshots, triggered line-scanner
The process interface (electronics within cable as well as industrial interface) must be
powered separately (5-24 VDC). Before switching on the power the PIF cable must be
connected to the camera.