5. Commissioning - Operation
Switching on the measuring device
By pressing the ON/OFF key, the Oxy-SMART is switched
on. The device heats up the sensor which has to be brought
to a preset temperature in order to allow a correct measuring
process. During this time it is not possible to perform any
measurements.
In the standby mode the cell is heated up to 50%.
Structure of the measuring equipment
Recommended structure:
Since most of purge gases are heavier than air, it is recommended to connect the
gas inlet at the lowest point of the system or the piping to expel the air and the
residual oxygen effectively.
Oxy-SMART
Weld joint
Gas outlet
Measuring
point
Gas inlet