10.
Checklist
of
safety
function
test
OptiFlex
-
K1
Safety
test
Protective earth conductor resistance
Ground leakkage current EN 60601 / IEC 601/ VDE 0751
Or
Ground leakkage current as in UL 2601
Function test
Measured
value
Date/
Signature
0,1 Ohm Ohm
500 µA µA
300 µA µA
OK Error
1.
Switch
on
the
OptiFlex
-
K1.
Press
the
two
outer control
(Extension/ Flexion)
simultaneously.
Software
version
VX.X
XX.XX.XX
(X
=
optional)
Keep
on
pressing.
Display: Artromot K1 Classic
2. The
maximum
range
of
motion
for
Extension/
Flexion
is
-
10
to
120
degrees.
Check
the
angle
in
position
0
degrees.
Tolerance
+/
-
5
degrees.
Check the angle in position 60 degrees. Tolerance +/- 5 degrees.
Check the angle in position 100 degrees. Tolerance +/- 5 degrees.
3. Check
the
emergency
-
off
function.
Start the OptiFlex K1 in any mode.
Press
any
control,
the
OptiFlex
-
K1
will
stop
immediately.
Check this for all controls.
5. Start
the
OptiFlex
-
K1
in
the
motion
range
between
10
to
120
degrees.
Set the speed to 100%.
Both
extre
me
points
should
be
reached
within
45
65
seconds.
13
4. Checking the "PAUSE" function.
Select the following settings:
Extension = 10 degrees
Flexion = 90 degrees
Speed = 50%
Pause = 10 seconds
Start the OptiFlex-K1.
At the reversal points (at 10 / 90 degrees) the pause must be lasting 10 seconds
(+/- 2 seconds).
The speed is substantially lower than in a 100% setting.