Protective earth conductor resistance
Ground leakkage current EN 60601 / IEC 601/ VDE 0751
Or
Ground leakkage current as in UL 2601
Function test
≤
≤
0,1 Ohm Ohm
500 µA µA
300 µA µA
OK Error
simultaneously.
Display: Artromot K1 Classic
Check the angle in position 60 degrees. Tolerance +/- 5 degrees.
Check the angle in position 100 degrees. Tolerance +/- 5 degrees.
Start the OptiFlex K1 in any mode.
Check this for all controls.
Set the speed to 100%.
4. Checking the "PAUSE" function.
Select the following settings:
Extension = 10 degrees
Flexion = 90 degrees
Speed = 50%
Pause = 10 seconds
Start the OptiFlex-K1.
At the reversal points (at 10 / 90 degrees) the pause must be lasting 10 seconds
(+/- 2 seconds).
The speed is substantially lower than in a 100% setting.