5
Motor
over
current:
The current for the motor exceeded the maximum limit
-
>
Check
the
mechanics
(Pos.
25)
-> Replace motor control (Pos. 20)
-> Replace the motor (Pos. 21)
6 Motor
control
error:
Internal error in the motor control.
-
>
Replace
motor
control
(Pos.
20)
7 Eprom
access
error:
Memory error in the access of the EPROM
-
>
Replace
the
hand
hel
d
programming
unit
(Pos.
14)
8 CPM
ROM
error:
Memory error in the motor control
-
>
Replace
motor
control
(Pos.
20)
9 Communication:
Communication to the motor control is not possible
-
>
Check
spiral
cable
and
connector (Pos. 15)
-> Replace the hand held programming unit (Pos. 14)
-> Replace motor control (Pos. 20)
10 General
error
in
the
motor
control:
Unknown error in the motor control
-
>
Replace
motor
control
(Pos.
20)
11 Motor
enable
timeout
Motor could not be enabled in time
-
>
Replace
motor
control
(Pos.
20)
12 Invalid
parameter
motor
error:
Motor has received a invalid parameter
-
>
Replace
motor
control
(Pos.
20)
-> Replace the hand held programming unit (Pos. 14)
13 Stop
release
error:
The motor could not be released
-
>
Replace
motor
control
(Pos.
20)
-> Replace the hand held programming unit (Pos. 14)
14 Unexpected
motor
Stop:
-
>
Check
spiral
cable
and
connector (Pos. 15)
-> Replace motor control (Pos. 20)
15 Motor
disabled:
Motor control disabled the motor.
-
>
Replace
motor
control
(Pos.
20)
16 Wrong
command
in
the
motor
:
-
>
Replace
motor
control
(Pos.
20)
-> Replace the hand held programming unit (Pos. 14)
17 5V
supply
error:
5V supply of motor control not sufficient
-
>
Replace
motor
control
(Pos.
20)
18 Initialise
error
real
time
clock:
-
>
Replace
the
hand
held
programming
unit
(Pos.
14)
19 Communication
error
real
time
clock:
-
>
Replace
the
hand
held
programming
unit
(Pos.
14)
20 Error
real
time
clock
:
-
>
Replace
the
hand
h
eld
programming
unit
(Pos.
14)
3
21
Range
exceeded:
The measured angle is out of the range of motion
-
>
Replace
motor
control
(Pos.
20)
22 ROM
error
in
the
hand
held
programming
unit:
Memory error in the hand held programming unit
-
>
Replace
t
he
hand
held
programming
unit
(Pos.
14)
23 Invalid
parameter:
Internal error in the hand held programming unit
-
>
Replace
the
hand
held
programming
unit
(Pos.
14)
24 24V
supply
error
motor
control:
Error in the 24V supply in the motor control
-
>
Replace
motor
control
(Pos.
20)
-> Replace the power supply electronics (Pos. 23)
25 Bus
error:
Bus system error
-
>
Replace
the
spiral
cable
of
the
hand
held
programming unit (Pos. 15)
-> Replace the hand held programming unit (Pos. 14)
-> Replace motor control (Pos. 20)
26 24V
supply
hand
held
programming
unit:
24V supply of the hand held programming unit is
defective
-
>
Replace
the
hand
held
programming
unit
(Pos.
14)
27 5V
supply
hand
held
programming
unit:
5V supply of the hand held programming unit is defective
-
>
Replace
the
hand
held
programming
unit
(Pos.
14)
28 3.3V
supply
hand
held
programming
unit:
3.3V supply of the hand held programming unit is
defective
-
>
Replace
the
hand
held
programm
ing
unit
(Pos.
14)
29 Calibration:
The calibration data in the motor control are wrong.
-
>
Perform
a
calibration
(see
chapter
8.1)
30 Calibration
error:
-
>
Repeat
the
calibration
(see
chapter
8.1)
-> Replace knee electronics (Pos. 3)
-> Replace motor control (Pos. 20)
31 Calibration
timeout
:
-
>
Replace
motor
control
(Pos.
20)
32 Motor
enable
error:
The motor could not be enabled
-
>
Replace
motor
control
(Pos.
20)
33 Motor
disable
error:
The motor could not be disabled
-
>
Re
place
motor
control
(Pos.
20)
34 Motor
stop
error:
Motor stop command timeout error
-
>
Replace
motor
control
(Pos.
20)
35 Configuration
error:
Invalid configuration of the hand held programming unit
-
>
Replace
the
hand
held
programming
unit
(Pos.
14)