The model SONOST 3000 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model SONOST 3000 should assure
that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV
contact
8 kV air
6 kV
contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power
supply lines
1 kV for
input/output
lines
2 kV for
power
supply lines
1 kV for
input/output
lines
Main power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV
differential
mode
2 kV
common
mode
1 kV
differential
mode
2 kV
common
mode
Main power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruption, and
voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip
in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in
U
T
)
for 5 cycles
70 % U
T
(30 % dip in
U
T
)
< 5 % U
T
(> 95 % dip
in U
T
)
for 5 s
< 5 % U
T
(> 95 % dip
in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in
U
T
)
for 5 cycles
70 % U
T
(30 % dip in
U
T
)
< 5 % U
T
(> 95 % dip
in U
T
)
for 5 s
Main power quality should be that of a
typical commercial or hospital
environment. If the user of the model
SONOST 3000 requires continued
operation during power mains
interruptions, it is recommended that the
model SONOST 3000 be powered from an
uninterruptible power supply or battery.