Owlet Dream Sock - Theory of Operation; Clinical SpO2 Accuracy Study Summary (continued)

Owlet Dream Sock
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80% to 90%
± 2.67 % Arms
± 2.22 % Arms
90% to 100%
± 2.46 % Arms
± 2.63 % Arms
Results of the clinical study demonstrate Dream Sock meets established criteria for SpO2 accuracy (in non-motion conditions) for the
range 70% to 100% SaO2 per the referenced standards.