8. If you are using an auto-LLT transfer siphon it is best to run it in manual mode. When the system is in auto mode the flow rate may
change too rapidly, and good stability may not be achieved.
9. The PID settings on the temperature controller may be different from those given in the test results. Typically, the P and I values
should be increased slightly.