EasyManua.ls Logo

pajunk MultiStim SWITCH - Page 33

pajunk MultiStim SWITCH
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English
43
d
.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
(according to EN 60601-1-2:2007; 5.2.2.2 Table 3)
MultiStim SWITCH is intended for use in the electromagnetic environment specifi ed below. The
customer or user of the MultiStim SWITCH should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic envi-
ronment - guidance
Conducted RF
according to
IEC 61000-4-6
3V
rms
150 kHz to 80 MHz
outside the ISM-bands
a
Not applicable Portable and mobile
RF communications
equipment should not
be used no closer to
any part of the Multi-
Stim SWITCH, includ-
ing cables, than the
recommended separa-
tion distance calculated
from the equation
applicable to frequency
of the transmitter.
Recommended separa-
tion distance:
d = 3,5/U1P for
150kHz to 80MHz
d = 12/U1P for
80MHz to 800MHz
d = 23/U1P for
800MHz to 2,5GHz
where P is the maxi-
mum output power
rating of the trans-
mitter in watts (W)
according to the trans-
mitter manufacturer
and d is the recom-
mended separation dis-
tance in meters (m)
b
.
10Vrms
150 kHz to 80 MHz in
the ISM-bands
a
Not applicable
Radiated RF
according to
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
10V/m Field strengths from
xes RF transmitters,
as determined by an
electromagnetic site
survey,
c
should be less
than the compliance
level in each frequency
range.
d
Interference may occur
in the vicinity of equip-
ment marked with the
following symbol:
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
XS190174B MultiStim SWITCH GB.indd 43XS190174B MultiStim SWITCH GB.indd 43 12.03.10 13:5312.03.10 13:53