How to fix Palm MicroBeam when the laser is not cutting?
- Aamy78Aug 3, 2025
If the laser is not cutting, you should: * Check energy and focus settings as described in the user manual, if necessary, take factory settings. * Check the red and green LEDs on the laser unit. If the red or green LED is off, check the key switch on the laser unit or the function of the safety switch in the illumination arm of the microscope, respectively. * Ensure the shutter is open. If the shutter is closed, open it. * Confirm that the filter revolver is switched to 'laser' (unless you have fluorescence equipment with P.A.L.M. simultaneous cutting). * Also, verify the position of the adjusting aid for the fluorescence lamp, the shutter for FL, both sliders with diaphragm, the filter slider, and the correction ring of the objective according to the thickness of the slide.