Paragon GUARDIAN - Output and Display Scaling Adjustments

Paragon GUARDIAN
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
6.2.2.2 VOLTAGE (Uni-polar Range)- Adjust R75 (ZERO) for a reading of
0.00vdc, .01vdc between J2 PROCESS (OUT+) and (GND) terminals.
VOLTAGE (Bi-polar Range)- Adjust R75 (ZERO) for a reading of 5.00vdc,
.01 vdc between J2 PROCESS (OUT+) and (GND) terminals.
6.2.3 With full scale pressure applied to the high port (See label on back of access door)
6.2.3.1 CURRENT - Adjust R87 (SPAN) for a reading of 20.0mA.
6.2.3.2 VOLTAGE - Adjust R84 (SPAN) for a reading of 10.00vdc between J2
(OUT+) and (GND) terminals.
6.2.3.3 Adjust R38 (DISPLAY) for full scale value on the LED display.
6.2.4 Repeat steps 2 and 3 to verify that they meet specification tolerances. Large
adjustments to either zero or span may result in a slight shift in the other adjustment.