■Avoid transporting over uneven floors.
This may cause damage to the bed, or deformation
or failure of the bed. If moving on uneven floor is
necessary, move the bed as slow as possible.
■The handheld controller must be attached to a suitable bed.
If it is attached to an unsuitable bed, the bed could operate unexpectedly, resulting in injury
or bed failure.
●Even though the bed is the same series, the suitable handheld controller may be different due to a
specification changes. Before attaching the handheld controller to the bed, check that the part number
and shape of the handheld controller are the same as those of the bed.
■When using a patient lifter, do not use the leg part height is more than 140mm.
The leg part of the patient lifter cannot be inserted under the bed and the patient lifter cannot be used
The angle of the leg section base must be adjusted by at least two persons from both
sides of the bed.
If only one person attempts to adjust the angle, the person may get caught between the leg section
base and main frame, resulting in injury.
■When using a patient lifter, it can be used in the area as shown in the picture.
●Do not use the patient lifter in the other area because it may damage or deform.