Abnormal ticking sound
occurred during back raise
operation.
Check if emergency CPR function
is operated previously.
Press the [] button of the back
raise adjustment buttons on the
handheld controller or the side-rail
integrated panel for more than 5
seconds in power on status.
19, 21,
24
The bed does not move
even when the handheld
controller buttons are
pressed.
Check if the bed was operated
continuously.
Reset the protect condition.
19
The bed does not move
even when the side-rail
integrated panel and the
patient control panel
buttons are pressed.
Check if the bed was operated
continuously.
Reset the protect condition.
22
Check if the Lock-out switch is
in the lock position.
Rotate the switch to the unlock
position.
20
The bed cannot be
moved.
Problem
Check
Corrective action
Refer-
ence
page
Check if the power plug is
plugged into the outlet.
Plug the power plug into the
outlet.
16
The bed does not move
even when the handheld
controller, the side-rail
integrated panel or the
patient control panel
buttons are pressed.
The power plug is plugged into
the outlet but the bed still does
not move.
Unplug the power plug from
the outlet and plug it in again.
Plug another electrical
device into the outlet to
check that power is available.
16
Check if the casters are locked.
Unlock the caster.
26-27
Check if the side-rail integrated
panel and the patient control
panel lock function is OFF.
Press the lock button to switch
the side-rail integrated panel and
the patient control panel lock
function OFF.
22
The bed cannot be
positioned correctly.
Check if there is nothing in the
moving parts on the bed.
Remove any obstacles.
-
Beeping alert sound occurred
when the handheld controller,
the side-rail integrated panel
or the patient control panel
buttons are pressed.
Check if the battery is charged.
Plug the power plug into the outlet.
34
The board will not come
off.
Check if the board is locked.
Unlock the board stoppers.
14
The bed does not move in
the intended direction
when the patient is being
transported.
Check if the caster is in the
steering position.
Unlock the caster in the
steering position.
26
51
Problems may not always indicate faults. Check the following once again before requesting repair.
If the problem persists after taking corrective actions, stop using the bed immediately, disconnect the
power plug from the outlet, and contact Paramount Bed or your distributor for repair.
Bed damaged in earthquakes, fires, floods or other incidents must be inspected and repaired by
Paramount Bed or your distributor. Otherwise, there is a risk of electric shock or fire due to
short circuit or current leakage from electrical parts, of injury due to malfunctioning caused by
deformation of the bed.
TROUBLESHOOTING