Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The A5 series is intended for use in the electromagnetic environments specified below.
The customer or the user of the A5 series must use it in an environment specified here.
The A5 Series is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
The A5 Series uses RF energy only for its internal functions.
Therefore, its RF emissions are low and are not likely to cause
any interference with nearby electronic equipment.
Electromagnetic Environment - Guidance
Compliance
Group 1
Class A
Class A
Complies
Voltage fluctuations / Flicker emissions
IEC 61000-3-3
Harmonic emissions
IEC 61000-3-2
RF emissions
CISPR 11
RF emissions
CISPR 11
Emission test
2
Handheld controller cable
1.5
Without shield
No.
Cable
Maximum length [m]
Remarks
1
AC power cable
4.6
Without shield
56
The A5 series requires special safety precautions regarding EMC to be observed. Install, operate, and
use the A5 series according to the information provided below.
The A5 series is intended for use in an electromagnetic environment specified according to
“Electromagnetic interference”.Using the A5 series in an unintended environment may result in the
occurrence of the following problems.
The function for supporting the patient’s body weight may not be maintained.
The emergency back-lowering function may not be maintained.
The A5 series is intended for use in doctor’s offices, clinics, exclusive medical facilities, independent
surgery centers, multiple treatment facilities, hospitals (except near high-frequency surgical equipment, and
outside RF shielded rooms for magnetic resonance images). The customers or users of the A5 series must
ensure that the A5 series is always used in thisenvironment.
Due to its emission characteristics, the A5 series is suitable for use in hospitals. If used in a home
environment, this device may not offer sufficient protection against radio frequency communication services.
It may be necessary for the user to take relaxation measures such as re-arranging or changing the direction of
the device.
If accessories, transducers, or cables other than those specified and sold by the manufacturers of the A5 series
are used as replacement parts for internal components, it may result in an increase in emission or a decline in
immunity of the A5 series.
Use portable RF communication devices (including antenna cables and peripheral devices such as external
antennas) at a distance of at least 30 cm from all parts of the A5 series. Using the RF communication
devices within a distance of 30 cm may cause a decline in the performance of the A5 series.
The A5 series cannot be used adjacent to another device or in a stacked state. If the A5 series must be used
adjacent to another device or in a stacked state, make sure the A5 series operates appropriately at the location of
use.
SPECIFICATIONS