Tab. 5.1 Regular Check Items
Items Content
Before
turning on
the power
Appearance
No smudge, rust, deformity, damage, looseness, distortion and
dampness.
Buttons
No damage or looseness.
Emergency release buon is in the pop-up reset state. You can
press this buon to release your arm in case your arm is locked in
the device because of sudden power loss.
LCD Screen No smudges or damage of the LCD external cover.
Connecon of
power supply
No damage of the AC adapter and power cable.
No heavy object hanging from the AC power cable.
The AC adapter plug is correctly fit in the device.
The AC adapter plug is correctly fit in the socket.
Numerical keypad No deformity, damage or looseness.
Thermal sensitive
printer
No deformity, damage or looseness.
No deformity or damage in the supporting mains charger or
RS232 connecting cable.
Cuff There is no smudging, deformity, damage, looseness or distortion.
After
turning on
the power
Appearance
No smoke or abnormal odor.
No abnormal sound or heat.
Buttons Button pressing is normal.
LCD Screen Display of measuring screen is normal.
Measuring All connectors are correctly connected.
Pressure Cuff pressuring-up and air release are normal.
Date/Time Setup of date and time are normal.
Calibration
Do not attempt to calibrate the device by yourself.
Calibration should be performed only by authorized service only. If you need this service, please
contact your local supplier.
※ If the device is used for 300,000 times, the LCD screen will flicker automatically (About 5 seconds).
Please contact your local supplier to return the device to the factory for recalibration.
Repair
Do not test or repair the device by yourself. Repair should be done only by customer service personnel
authorized by the manufacturer.
Disposal
Please dispose of and recover the device under the direction of the local environmental protection
authority based on environmental protection principles.
※ After the use period of the product of 3 years expires, it should be returned to the factory for
performance testing, and it can be used further if all indexes are to the standard; the company will