EasyManuals Logo

PCE Instruments PCE-322A Instruction Manual

PCE Instruments PCE-322A
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
21
AC: Output voltage: 1Vrms corresponding to each range
step.
Output impedance: 100Ω
DC: Output voltage: 10mV/dB
Output impedance: 1kΩ
6. CALIBRATION PROCEDURES
Make the following switch settings:
Frequency weighting: A-weighting
Time weighting: FAST
Level range: 50 ~100dB
Insert the microphone housing carefully into the 1/2
inch insertion hole of the calibrator94dB @ 1kHZ.
Turn on the switch of calibrator and adjust the CALL
potentiometer of the unit 94.0dB is displayed.
NOTE: All products are well calibrated before shipment.
Recommended recalibration cycle: 1 year.

Other manuals for PCE Instruments PCE-322A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PCE Instruments PCE-322A and is the answer not in the manual?

PCE Instruments PCE-322A Specifications

General IconGeneral
BrandPCE Instruments
ModelPCE-322A
CategoryData Loggers
LanguageEnglish

Related product manuals