7
WARNING: Handle lithium batteries
carefully, observe warnings on battery casing.
Dispose of in accordance with local regulations.
RECONDITION PROCEDURE
Exposure of the internal sensor to chemical vapours
may interfere with the internal sensor and cause
inaccurate readings to be logged. In a clean
environment, this will slowly rectify itself. However,
exposure to extreme conditions or chemical vapours
will require the following reconditioning procedure to
bring the internal sensor back to calibration state.
80°C (176°F) at<5%RH for 36h (baking) followed by
20-30°C (70-90°F) at>74%RH for 48h (re-hydration)
High levels of pollutants may cause permanent
damage to the internal sensor.
1
2
3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PCE Instruments PCE-HT 71N and is the answer not in the manual?

PCE Instruments PCE-HT 71N Specifications

General IconGeneral
Accuracy Temperature±0.5 °C
Resolution Temperature0.1 °C
Resolution Humidity0.1% RH
Memory16, 000 readings
PC interfaceUSB
SoftwarePCE-HT 71N software
TypeData Logger
Measurement Range Temperature-40 °C to +70 °C
Measurement Range Humidity0 %RH to 100 %RH
Accuracy Humidity±3 %RH
Sampling rate1 sec to 24 hrs
Battery lifeApprox. 1 year
Operating conditions-40 °C to +70 °C, 0 %RH to 100 %RH
Storage conditions-40 °C to +70 °C, 0 %RH to 100 %RH
SensorThermistor and capacitive humidity sensor