6
* To save power, flashing-Cycle can be changed to 20s or 30s
via analysis-software. For more details, please see the guide
of analysis-software.
** To save power, all alarm indications for temperature and
humidity can be disabled. For more details, please
see the guide of analysis-software.
***When both temperature and relative humidity readings
exceed alarm level synchronously, LED status
indication alternate every cycle.
****When battery low, all operations will be disabled
automatically. NOTE: No matter what have data-logging
interrupted any time, you must use the analysis-software to
restart logging.
BATTERY REPLACEMENT
We recommend that you replace the battery every 12
months, or prior to logging critical data.
Datalogger does not lose its stored readings when the
battery is discharged or when the battery is replaced;
the data logging process will however be stopped and
cannot be re-started until the battery has been
replaced and the logged data has been downloaded to
PC.
Only use 3.6V lithium batteries. Before replacing the
battery, remove the model from the PC.
NOTE:
Leaving the model plugged into the USB port for
longer than necessary will cause some of the battery
capacity to be lost.