8. Appendix
8.1. Oscilloscope
Only if another instructions are provided, are all technical specifications applicable to the probe with the 10X attenuation
switch setting and the PeakTech
®
1220 digital oscilloscope. In order to be up to these specifications, the oscilloscope
should meet the following requirement.
The instrument should operate continuously for more than 30 minutes under the specified operating temperature.
If the operating temperature range of variation is up to or larger than 5 Celsius degrees, the system function menu must
be opened to make the system perform a self- calibration” procedure.
Except those specifications marked with the word Typical, all specifications can be up to.
Sampling
Sampling modes
Normal sampling
Peak detection
Average value
Sampling rate
500 MSa/s
Input
Input coupling
DC, AC,Groud
Input impedance
1M±2% connected in parallel with 20pF±5pF
Probe attenuation coefficient
1X, 10X, 100X, 1000X
Max. Input voltage
400V (peak)
Channel delay time (typical)
150 ps
Horizontal
Sampling rate range
0.25S/s500MS/s
Waveform interpolation
(sin x)/x
Record length
6K points
Scanning speed rangeS/div
5ns/div100s/div, stepping in the “1-2-5 mode.
Time interval ( T)measurement accuracy (full
bandwidth)
Single:
±(1 sampling interval time+100ppm×reading+0.6ns)
>average 16:
±(1 sampling interval time +100ppm×reading+0.4ns)
Vertical
Analog digital converter (A/D)
Resolution of 8 bits
Sensitivity range (V/div)
5mV/div5V/divat the input BNC
Displacement range
±2 V (5 mV/div 200 mV/div)
Analog bandwidth
20 MHz
Single bandwidth
Full bandwidth
Low frequency response
(AD coupling, -3dB)
10 Hz (at the BNC)
Rise time (typical one at the BNC)
17.5ns
DC gain accuracy
±3%
DC measurement accuracy (average value
sampling mode)
The voltage difference (V) between any two points on the
waveform after averaging the captured waveforms more
than16: ±(5% reading + 0.05 divisions).
-121-