8. Appendix
8.1. Oscilloscope
Only if another instructions are provided, are all technical specifications applicable to the probe with the 10X attenuation
switch setting and the PeakTech
®
1220 digital oscilloscope. In order to be up to these specifications, the oscilloscope
should meet the following requirement.
The instrument should operate continuously for more than 30 minutes under the specified operating temperature.
If the operating temperature range of variation is up to or larger than 5 Celsius degrees, the system function menu must
be opened to make the system perform a “self- calibration” procedure.
Except those specifications marked with the word Typical, all specifications can be up to.
Sampling
Normal sampling
Peak detection
Average value
1M±2% connected in parallel with 20pF±5pF
Probe attenuation coefficient
Channel delay time (typical)
Scanning speed range(S/div)
5ns/div~100s/div, stepping in the “1-2-5” mode.
Time interval ( △ T)measurement accuracy (full
bandwidth)
Single:
±(1 sampling interval time+100ppm×reading+0.6ns)
>average 16:
±(1 sampling interval time +100ppm×reading+0.4ns)
Analog digital converter (A/D)
Sensitivity range (V/div)
5mV/div~5V/div(at the input BNC)
±2 V (5 mV/div – 200 mV/div)
Low frequency response
(AD coupling, -3dB)
Rise time (typical one at the BNC)
DC measurement accuracy (average value
sampling mode)
The voltage difference (△V) between any two points on the
waveform after averaging the captured waveforms more
than16: ±(5% reading + 0.05 divisions).