Questions and Answers

  • B
    Brittany AlvaradoAug 18, 2025
    What to do if Pedigo Elite Warming Drawer displays error E-10, ES10, ES20, ES30, ES40, ES50, ES60, or ES70?
    • J
      Jonathan DelacruzAug 18, 2025
      If your Pedigo Warming Drawer shows error E-10, ES10, ES20, ES30, ES40, ES50, ES60, or ES70, it indicates the sensor is shorted. The software will disengage the heating pads, and you should acknowledge the error by pressing the ON/OFF button. If the error continues, a qualified technician should test the sensor. To test the sensor, detach it from the unit and use an Ohm meter to measure its resistance. Check the sensor at 0°C (32°F) using ice water. A reading of 32.6 KOhm ±1.5 KOhm suggests replacing the display, while a reading of ±2 KOhm indicates replacing the sensor. Also, check the wires for integrity and ensure secure connections at the control and terminal block. Resecure any faulty connections if necessary. If the error persists after these steps, contact a service technician.
  • K
    Kimberly GalvanAug 20, 2025
    What does error E-11, ES11, ES21, ES31, ES41, ES51, ES61, or ES71 mean on my Pedigo Warming Drawer?
    • H
      Hayden HansenAug 20, 2025
      If your Pedigo Warming Drawer displays error E-11, ES11, ES21, ES31, ES41, ES51, ES61, or ES71, it means the sensor is open. The software will turn off the heating pads. You should press the ON/OFF button to acknowledge the error. If it continues, have a qualified technician test the sensor. To test it: remove the sensor from the unit, use an Ohm meter to check its resistance, and test it at 0°C (32°F) in ice water. If the reading is 32.6 KOhm ±1.5 KOhm, replace the display. If it reads ±2 KOhm, replace the sensor. Also, inspect the wires for damage and ensure the connections at the control and terminal block are secure. If the error still occurs, call a service technician.
  • C
    Cassandra ThompsonAug 22, 2025
    What does it mean when my Pedigo Warming Drawer shows error P130, P230, P330, P430, P530, P630, or P730?
    • A
      Amanda EspinozaAug 23, 2025
      If your Pedigo Warming Drawer shows error P130, P230, P330, P430, P530, P630, or P730, it indicates that the heating pad has not reached the set-point temperature. You should acknowledge the error by pressing the ON/OFF button. If the error persists, a qualified service technician should test the heater pad(s). To test the heater pad(s): * Turn the unit OFF and unplug it from AC power. * Use an Ohm meter to measure the resistance between the L (Line) and N (Neutral) leads of a cold heater pad. The Ohm readings should be: 120V = 72 Ohm ±10% or 230V = 288 Ohm ±10%. If the error continues after these steps, contact a service technician.
  • R
    Ronnie TorresAug 25, 2025
    What to do if Pedigo Warming Drawer displays error P131, P231, P331, P431, P531, P631, or P731?
    • K
      Kristen McclureAug 26, 2025
      If your Pedigo Warming Drawer shows error P131, P231, P331, P431, P531, P631, or P731, it means there is a heater pad over-temp error. The software will disengage the heating pads. You should first acknowledge the error by pressing the ON/OFF button and then allow the unit to cool down. If the error persists after cooling, contact a service technician.
  • C
    Carolyn MillerAug 28, 2025
    What does error *E-98 mean on a Pedigo Elite?
    • H
      harrisbrianAug 28, 2025
      Error *E-98 on your Pedigo Warming Drawer indicates that the temperature difference between cavity sensors 1 and 2 is more than 1.7ºC (3ºF). To resolve this: * Remove the product and allow the unit to cool down. * Verify that the product sensor is clean and operating correctly. * Press the power button to clear the error code. If the error continues, a qualified service technician should replace the sensor switch assembly. Contact service.
  • B
    Brendan CowanAug 30, 2025
    What does error *EFAn mean on a Pedigo Elite Warming Drawer?
    • M
      Mr. Stephen HendersonAug 30, 2025
      Error *EFAn on a Pedigo Warming Drawer (Fluid Warmers Only) indicates a fan or fan sensor failure. If the fan is running, make sure the sensor lens is clean. If the error persists, use an Ohm meter to check the sensor wires; a good wire will read less than 1 Ohm. If the wires are good, replace the sensor. If the error continues after replacement, check the sensor mount location. If the error still persists, replace the control.
  • D
    Donna LeeSep 1, 2025
    How to fix error E-90 on Pedigo Elite Warming Drawer?
    • T
      Thomas WeaverSep 2, 2025
      Error E-90 on your Pedigo Warming Drawer means that a button has been held down for more than 60 seconds. To fix this, adjust the control. The error will reset once the problem is resolved.
  • G
    Glenn CampbellSep 4, 2025
    What does E-95 error mean on Pedigo Elite Warming Drawer and how to resolve it?
    • A
      amandareidSep 4, 2025
      Error E-95 on your Pedigo Warming Drawer indicates a factory test pin short has been detected. To resolve this, first ensure that there is no debris causing a short between the test pins. If the pins are clear and the error persists, the control should be replaced.
  • M
    munoznicholasSep 6, 2025
    What does E-60 mean on my Pedigo Warming Drawer?
    • M
      Melissa LaneSep 7, 2025
      Error E-60 on Pedigo Warming Drawers (Blanket Warmers Only) indicates that the Real Time Clock rechargeable battery backup has discharged. To resolve this, plug the unit into an outlet for 30 minutes. Then, refer to the “Timer Control Panel” section in the blanket manual to reset the clock.
  • M
    Mr. Chad HansenSep 9, 2025
    What does error *E-70 mean on my Pedigo Elite?
    • T
      Tyler MoranSep 9, 2025
      Error *E-70 on your Pedigo Warming Drawer means that more heater pads are detected than are set for. To correct this, hold the ON/OFF button for 12 seconds until the display shows "PAd#” (# = number of pads selected [3-7]). Then, press the UP or DOWN arrow to adjust to the correct number of pads. Note that Fluid Warmers (2.5ft3 & 4.0ft3) = 3 pads, and Blanket Warmers (1.5ft3, 2.5ft3 = 3 pads, 3.5ft3, 4.0ft3, 7.5ft3 = 4 pads).