Operators Manual
Blanket Warmer
P-2040
P-2055
MN-39925
REV.02
8/20
EN

Other manuals for Pedigo P-2040

Questions and Answers

  • B
    bryan32Aug 9, 2025
    What to do if Pedigo Warming Drawer shows E-60 Real-Time Clock Checksum Error?
    • C
      Christina DonovanAug 9, 2025
      If your Pedigo Warming Drawer displays an E-60 error, it might be because the unit was unplugged for a while. Try this: turn the circuit breaker switch to the ON position for 1 minute, then turn it OFF for 5 seconds, and back to ON. The error message should disappear. After resetting, leave the unit plugged in with the circuit breaker ON for at least 24 hours to fully recharge. Also, verify that the date and time are correct.
  • S
    Sarah LewisAug 14, 2025
    Why is my Pedigo Warming Drawer showing E-30 Under Temp Error?
    • L
      Laura CarsonAug 14, 2025
      The E-30 error on your Pedigo Warming Drawer indicates that the blanket chamber temperature has been lower than the set temperature for 90 minutes or longer. First, check that the door is fully closed. Next, verify that the high limit switch at the rear of the unit hasn't tripped. If it has, reset it by pressing the limit reset button. Correct the cause of the trip. If the high limit switch won't reset, it's likely defective and needs replacement.
  • S
    Sean PattersonAug 25, 2025
    How to fix E-10 Sensor Short error on Pedigo P-2040?
    • M
      Melissa MunozAug 25, 2025
      Detach the sensor from the terminal block. Use an Ohm meter to measure the resistance of the sensor. Check sensor at 32°F (0°C) using a container of ice water. If Ohm reading is 100, replace display. If Ohm reading is ±10, replace sensor. Check wires for integrity and proper connections at the control and terminal block. If necessary, re-secure the faulty connections. If the error continues, further assistance may be needed.
  • P
    plopezAug 29, 2025
    What causes E-11 Sensor Open error on Pedigo P-2040?
    • S
      Samuel WilliamsAug 29, 2025
      Detach the sensor from the terminal block. Use an Ohm meter to measure the resistance of the sensor. Check sensor at 32°F (0°C) using a container of ice water. If Ohm reading is 100, replace display. If Ohm reading is ±10, replace sensor. Check wires for integrity. Check for proper and secure connections at the control and terminal block. If necessary, re-secure the faulty connections. If the error continues, further assistance may be needed.
  • S
    Stacey SimonSep 2, 2025
    What does E-99 Hardware Over Temp mean on Pedigo P-2040 Warming Drawer?
    • A
      Alexander DelacruzSep 3, 2025
      If your Pedigo Warming Drawer displays an E-99 error, inspect the connections and condition of the high limit bimetal thermostat. If the error continues, further assistance may be needed.
  • S
    Stephanie BellSep 10, 2025
    What to do if Pedigo P-2040 Warming Drawer displays E-31 Over Temp error?
    • R
      Robert WilliamsSep 12, 2025
      If your Pedigo Warming Drawer is showing an E-31 error, the unit may be overloaded. Try redistributing the inventory, making sure not to exceed the height of the basket assembly. Also, check the sensor at 32°F (0°C) using ice water; the sensor reading should be around 100 ohms. In some cases, the relay (solid state) or the control's temperature circuit may be defective. If the error persists, further assistance may be needed.