Fig. 3-6 The menu choices after selecting FILTER.
Analog Lowpass Filter
The counter has analog LP filters of RC type, one in each of the channels A and B, with a cutoff frequency of
approximately 100 kHz, and a signal rejection of 20 dB at 1 MHz.
Accurate frequency measurements of noisy LF signals (up to 200 kHz) can be made when the noise components have
significantly higher frequencies than the fundamental signal.
Digital Lowpass Filter
The digital LP filter utilizes the Hold-Off function described below.
With trigger Hold-Off it is possible to insert a deadtime in the input trigger circuit. This means that the input of the
counter ignores all hysteresis band crossings by the input signal during a preset time after the first trigger event.
When you set the Hold-Off time to approx. 75% of the cycle time of the signal, erroneous triggering is inhibited around
the point where the input signal returns through the hysteresis band. When the signal reaches the trigger point of the
next cycle, the set Hold-Off time has elapsed and a new and correct trigger will be initiated. Instead of letting you
calculate a suitable Hold-Off time, the counter will do the job for you by converting the filter cutoff frequency you enter
via the value input menu below to an equivalent Hold-Off time.
Fig. 3-7 Value input menu for setting the cutoff frequency of the digital filter.
You should be aware of a few limitations to be able to use the digital filter feature effectively and unambiguously. First
you must have a rough idea of the frequency to be measured. A cutoff frequency that is too low might give a perfectly
stable reading that is too low. In such a case, triggering occurs only on every 2nd, 3rd or 4th cycle. A cutoff frequency
that is too
high (>2 times the input frequency) also leads to a stable reading. Here one noise pulse is counted for each half-cycle.
43 / 50

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals