5. POST REPROCESSING
1) Following reprocessing, the endoscope may either be reused or placed in storage.
2) Prior to reuse, ensure that instrument has been properly inspected and fully prepared for the next clinical pro-
cedure.
3) Prior to storage, ensure that all internal channels, scope components, instrument surfaces and accessories are
thoroughly dry.
4) A cotton tipped applicator moistened with 70-90% ethyl or isopropyl alcohol for disinfection may be used to
carefully remove any films or residues left upon the lens surfaces, such as the distal objective lens.
5) The endoscope should be hung in a clean, dry, well-ventilated storage cabinet at room temperature. The
insertion tube and light guide cable should be hung and kept as straight as possible during storage.
CAUTION:
When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should
be allowed to return to room temperature prior to use and/or further handling.
CAUTION:
Make sure that all removable components such as the air/water valve, suction valve, rubber inlet
seal(s), and rubber distal tip are detached from the scope. This will allow for better air circulation
throughout the internal channels to ensure thorough drying.
Never store the endoscope, its components and accessories in the carrying case, as this type of dark,
humid and unventilated environment is conducive to bacteria colonization, which increases the risk of
cross contamination. These cases are intended for transportation of the instrument, not storage.
Never store the endoscope in areas of high humidity, high temperatures or in direct exposure to
sunlight or X-rays.
Avoid storage of the scope in cabinets, which have any sharp edges, exposed nails/screws, etc.
Contact with sharp objects can puncture, scratch or otherwise damage the endoscope.