1-3. Video Processor
NOTE:
Read the instructions for use supplied with the video processor.
WARNING:
Do not install, operate or store electro-medical equipment in a dusty
environment. Accumulation of dust within these units may cause
malfunction, smoke, or ignition.
NOTE:
Be sure to use compatible bottle and water bottle cap. If incompatible
bottle and a water bottle cap are used together, it could cause the air to
escape resulting in insufcient pressure and ow of air and water during
the endoscopic procedure.
1) PENTAX Video Processor
(6)
(1)
(2)
(5)
(4)
(3)
(1) Lamp Switch
(2) Pump Switch
(3) Endoscope Electrical
Connector
(4) Light Guide
Receptacle
(5) Power Switch
(6) Water Bottle Assembly
Figure 1.2
CAUTION:
Replace the lamp before the lamp life expires. Prior to use, check the lamp
life indicator. Excessive use of the lamp beyond the lamp life could cause
the lamp to explode resulting in damage to the video processor. Refer to
the video processor's instructions for use regarding the lamp life.
NOTE:
Software update may be required depending on the software version of
the PENTAX video processor. If the software is not updated, the image will
not be displayed. If the images are not displayed correctly, please contact
your local PENTAX service facility.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pentax EG27-i10 and is the answer not in the manual?

Pentax EG27-i10 Specifications

General IconGeneral
BrandPentax
ModelEG27-i10
CategoryMedical Equipment
LanguageEnglish

Related product manuals