User’s Manual Permobil M5 Corpus Troubleshooting
Event
Possible cause Remedy
The wheelchair switches itself off after a
certain period of inactivity (20-30 min).
The electronics’ energy saving mode has
been activated.
Switch the wheelchair on again using the
start button on the control panel.
The wheelchair stops while being driven. The cable connection to the control
panel has come loose.
Insert the cable in the control panel.
Main circuit breaker tripped.
See page 243.
The wheelchair can only be driven at
reduced speed.
Seat lift or seat angle raised too high.
Applies only to an powered seat lift and
seat angle.
Lower the seat lift or seat angle.
The wheelchair cannot be charged. Main circuit breaker switched to off
position after e.g. battery replacement.
See page 243.
The charging circuit breaker has tripped. Wait five minutes, the circuit breaker will
automatically reset.
Diagnostics R-net LCD control panel
General
When an error or a fault occurs in the wheelchair’s electronics,
information about it is displayed in the control panel display. This
information can then be used to diagnose where the error, or fault,
occurred and its cause.