16
3.4 System Set
This mode allows the user to set the working parameters, such as tungsten lamp On/Off, deuterium
lamp On/Off, work mode, time set, wavelength correction, and dark current correction.
In the case of main menu of the instrument, press ǏƷǐ and Ǐǐto move the cursor to the option
of system set. Then press ǏENTERǐ.
(Fig. 3-12)
(Fig. 3-13)
(Fig. 3-14)
Ĵ
Deuterium Lamp
In the case of system set, press ǏƷǐ and Ǐǐ to move the cursor to the D2 Lamp. Each time you
pressǏENTERǐ, the ON and OFF will change alternately. It costs 20 seconds for the deuterium lamp to
change from off to on. The system will display BUSY and neglect the user’s input during this period.
Ĵ
Tung st en Lam p
In the case of system set, press ǏƷǐ and Ǐǐ to move the cursor to the W Lamp. Each time you
pressǏENTERǐ, the ON and OFF will change alternately.
Ĵ
Work Mode
In the case of system set, press ǏƷǐ and Ǐǐ to move the cursor to the option of Work Mode. Each
time you press ǏENTERǐ, the MCU and PC will change alternately. The default work mode is MCU.
When the user selects PC mode, it will display the waiting page of PC connection, as following Fig (3-15).
The user should confirm the main unit have connected with PC properly, and then use PC software to