25
Set the change lamp wavelength, the instrument will switch over the lamp automatically.
Why isn’t the stray light eligible?
The poor power and the resultant raise of the dark current adversely affect it. Perform dark current
correction will solve the problem. In addition, if the instrument is placed in the humid environment for a
long time, it will also cause the stray light raises higher.
It is recommended that switch on the power of the instrument once a week to avoid some problems
caused by leaving unused.
How to clear the data?
T60 is designed to clear all the data after printout. The system will promote the user to print data after
changing the parameters. If the user selects “Don’t print”, the system will also clear all the data.