4
The MET portable hardness testers are designed to measure hardness of metals and alloys in the most
popular hardness scales, such as Rockwell C (HRC), Rockwell B (HRB), Brinell (HB), Vickers (HV), and
Leeb (LD). Calibration on these scales are carried out directly by Phase II prior to shipment of each unit.
The MET hardness testers offer additional “open” scales (H0, H1, H2, H3) that provide the opportunity
to:
Calibrate and conduct hardness tests in other scales (i.e., Superficial Rockwell (HRN and HRT)
or any other common hardness scale)
Conduct hardness test of metals that have significantly different properties from steel (iron,
aluminum and copper alloys etc.).
Setting Test Parameters
1) Scale Selection (Press Left or Right Arrow for selections)
<HRC> <HRB> <HB> <HV> <HLD> <H1> <H2> <H3>
Ø= Yes ×= No
2) Operation Mode Selection (Press Left or Right Arrow for selections)
<Test> <Calibration> <Processing>
Ø× Ø× Ø×
Ø× Ø×
Measurement
and Archive
Measurement on
Test Block #1
Sample limits setting
and average value
determination
Archive Scale Correction
Measurement on
Test Block #2
Scale Correction