1-2
PRODUCT INFORMATION
The unit incorporates an overcurrent trip circuit which will instantly de-energize the high voltage
power supply if such a condition is encountered. The level of sensitivity at which trip-out will occur
is controlled by the front panel adjustment labeled Min-Max. The trip setting can be as low as
10% or as high as approximately 110% of full scale reading. Overcurrent trip is indicated by an
audible alarm.
SPECIFICATIONS
Description:
DC High Potential Tester with Megohmmeter
Primary Application:
Dielectric withstand testing and insulation resistance measurement of motors, cables, and
transformers.
Input:
120 Volts AC, 50/60 Hz, single phase, .3 Amps
or
220 Volts AC, 50/60 Hz, single phase, .15 Amps
Output:
0-6 kV DC, continuously adjustable with manual control
Negative Output, Positive Ground
Voltmeter:
2%, 3.5, analog meter, 0-3/6 kV
Currentmeter/Megohmmeter:
2%, 3.5, analog meter,
0-.5/5/50/500/5000 microamperes DC
0-.5-50/500/5000/50,000/500,000 megohms at 2.5 kV DC
1-100/1000/10,000/100,000/1,000,000 megohms at 5 kV DC
Termination:
Input: Standard modular 3-wire cord mating to fused input module
High Voltage: 10 attached cable with clip and insulator
Return: 10 separate cable with insulated clip (metered and guard)
Ground: 10 separate cable with clip
Dimensions:
18 W x 11 D x 11.5 H (457 mm W x 279 mm D x 292 mm H)
Weight:
21 lbs. (9.5 kg) 120 VAC
22 lbs. (10 kg) 220 VAC