EasyManua.ls Logo

Phenix PM6 - Installation and Operation Instructions; Installation; Operation

Phenix PM6
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-1
INSTALLATION AND OPERATION INSTRUCTIONS
INSTALLATION
Install the tester as follows:
1. Turn the main Power switch off (out position, non-illuminated).
2. Turn the High Voltage power switch off (out position, non-illuminated).
3. Connect the AC Line cord to a 115 Volt source.
4. Make necessary grounding connections.
5. Set all switches and controls present on your tester as listed below for initial start-up:
Raise Voltage 0 (zero)
Ammeter/Megohm Range X1/5000
Ground Jumper Jumped to Return or Guard as required by test
KV Meter 6 kV
Current Trip Mid-range
6. Connect the test leads to the specimen to be tested. See Test Lead Connections section for
information on how to connect the test leads.
OPERATION
1. Turn the unit on by pressing the Power switch. The switch will illuminate and remain
recessed.
2. Energize the High Voltage power supply by pressing the High Voltage power switch. The
switch will illuminate and remain recessed.
NOTE: If the Raise Voltage control has not been set to zero, the overcurrent alarm will
sound. To silence this alarm insure that the raise voltage control is in its full counter-
clockwise position.
3. Rotate the Raise Voltage control clockwise until the desired test voltage is observed on the
Kilovolt meter. Maintain output voltage at the desired level for the required amount of time.
If the test was initiated on the 6 scale and a more accurate voltage reading is required (3 kV
or below), select 3 on the KV Meter range switch.
NOTE: If an overcurrent condition is encountered before required test voltage is reached and
it is desirable to achieve a higher test voltage, adjust the Min-Max to a more clockwise
position.

Related product manuals