Chapter 3 Ventilator Alarms
25
Check AC Power Supply
This info message occurs when the AC power input to the ventilator is incorrect. The device
continues to operate but therapy may not start.
External Battery Disconnected
This info message occurs when an external battery is disconnected from the ventilator while
operating. The device continues to operate on AC power.
Battery Discharging Stopped Due to Temperature
This info message occurs when the detachable battery becomes overheated while providing power
for the device. The device continues to operate. The battery is not used and the power source is
switched to the next available power source.
Battery Not Charging Due to Temperature
This info message occurs when the detachable battery becomes too hot while charging or the
device was in too cold or hot an environment before charging started. The device continues to
operate. Battery charging stops until the battery cools or warms suciently.
Battery Not Charging
This info message occurs when the device has detected a condition that prevents the battery from
accepting a charge. The device continues to operate. Battery charging stops.
External Battery Depleted
This info message occurs when the external battery is fully depleted. The device continues to
operate using the detachable battery if it is available.
Detachable Battery Disconnected
This info message occurs when the detachable battery power source is lost and the device has
switched to an alternate power source. If detachable battery power returns, the ventilator will beep,
but no message will appear on the display.