BiPAP A40 user manual
30
Message Priority Audible Visual Indicators Device
Action
User Action
Start On Battery Info
◊
Start On Battery
message
Operates Check battery status. Connect
to AC power source as soon as
possible.
Check AC
Power Supply
Info
◊
Check AC Power
Supply message
Operates Connect ventilator to a
battery and remove AC power.
Replace the AC power supply.
If condition persists, contact
your home care service
provider.
External Battery
Disconnected
Info
◊
External Batt
Disconnected
message
Operates Check the connection to the
battery if not an intentional
disconnection.
External Battery
Depleted
Info
◊
Ext Batt Depleted
message
Operates Replace depleted external
battery with another or switch
to AC power, if available.
Recharge depleted external
battery.
Battery
Discharging
Stopped Due to
Temperature
Info
◊
Batt Discharge
Stopped - Temp
message
Operates Move device to cooler
location. Make sure device
is not close to a heat source.
Make sure cooling vents are
not blocked. If condition
persists, contact your home
care service provider.
Battery Not
Charging Due to
Temperature
Info
◊
Batt Not Charging -
Temp message
Operates Make sure device is not
close to a heat source. Make
sure cooling vents are not
blocked. Move device to a
cooler location. If device is
too cold, allow it to warm up.
If condition persists, contact
your home care service
provider.
Battery Not
Charging
Info
◊
Detach Battery Not
Charging message
Operates Replace the battery or nd
an alternate power source. If
condition continues, contact
your home care service
provider.
Detachable
Battery
Disconnected
Info
◊
Det Batt Disconnected
message, and a box
appears around
battery in use.
Switches to
alternate
power
source
Check connection of the
detachable battery to
ventilator. Check the charge
available on detachable
battery and recharge battery if
necessary.