Chapter 5 Viewing and Changing Settings
51
Low Tidal Volume Alarm
Select On to enable or O to disable the Low Tidal Volume alarm. When the alarm is enabled,
an audible indicator sounds if target tidal volume can’t be reached. This alarm is only available
when AVAPS is enabled (or in AVAPS-AE mode).
Low Minute Ventilation Alarm
This setting enables or disables the Low Minute Ventilation alarm. The alarm activates when the
calculated minute ventilation is less than or equal to this setting. Select O to disable this alarm,
or increase or decrease the setting from 1 l/min to 99 l/min in increments of 1.
High Respiratory Rate Alarm
This setting enables or disables the High Respiratory Rate alarm. The alarm activates when the
measured respiratory rate reaches or exceeds this setting. Select O to disable this alarm, or
increase or decrease the setting from 1 BPM to 60 BPM in increments of 1.
5.6.2 Changing Options Menu Settings
1. From the Main Menu screen, use the Up/Down key to highlight the Options item.
2. Press the Right key to select Options.
Options Settings
The following settings are available on the Options menu.
Menu Access
Select Full or Limited menu access. Full menu access allows home care service providers to
access all ventilator and prescription settings. Limited menu access allows users to access only
certain settings and does not allow them to change prescription settings.
Detailed View
Turn Detailed View On or O using this setting. Detailed view displays additional therapy
information on the Monitor screen.
Language
Select the Language that the software will appear in (English, French, German, etc.). The
information on the screens will display in the language selected here.