Chapter 6 Cleaning and Maintenance
59
6.1 Cleaning the Ventilator
The ventilator’s exterior surface and the exterior of the detachable battery pack compartment and
battery pack (if using) should be cleaned before and after each patient use and more often if needed.
1. Unplug the device and clean the front panel and exterior of the enclosure as needed using a
clean cloth dampened with water and a mild detergent.
2. Inspect the device and tubing for damage after cleaning. Replace any damaged parts.
3. Allow the device to dry completely before plugging in the power cord.
6.1.1 Cleaning for Multiple Users
Warning: If you are using the device on multiple users, discard and replace the bacteria lter each time the device is
used on a dierent person.
When using the device on multiple users, complete the following steps to clean the device before
each new user.
1. Unplug the device before cleaning.
2. Clean the outside of the device only. Use a cloth with one of the following cleaning agents to
clean the exterior of the device:
- Mild detergent
- Hydrogen Peroxide, 3%
- 91% Isopropyl Alcohol
- 70% Isopropyl Alcohol
- Vinegar, 5% acidity
- Water
- Chlorine bleach, household, 5.25% sodium hypochlorite, 1 to 5 part reduction with water
- DisCide Towelettes
3. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.
4. Allow the device to dry completely before plugging in the power cord.
6. Cleaning and Maintenance