6.2 Cleaning and Replacing the Air Inlet Filter
Under normal usage, you should clean the gray foam lter at least once every two weeks and
replace it with a new one every six months. The white ultra-ne lter is disposable and should be
replaced after 30 nights of use or sooner if it appears dirty. DO NOT clean the ultra-ne lter.
1. If the device is operating, stop the airow. Disconnect the device from the power source.
2. Remove the lter(s) from the enclosure by gently squeezing the lter in the center and pulling
it away from the device.
3. Examine the lter(s) for cleanliness and integrity.
4. Wash the gray foam lter in warm water with a mild detergent. Rinse thoroughly to remove all
detergent residue.
5. Allow the lter to air dry completely before reinstalling it. If the foam lter is torn or damaged,
replace it. Only Philips Respironics-supplied lters should be used as replacement lters.
6. If the white ultra-ne lter is dirty or torn, replace it.
7. Reinstall the lters, inserting the white ultra-ne lter rst if applicable.
6.3 Cleaning the Reusable Tubing
1. Clean the reusable tubing before rst use and daily.
2. Disconnect the exible tubing from the device.
3. Gently wash the tubing in a solution of warm water and a mild detergent.
4. Rinse thoroughly and air dry.
5. Inspect the tubing for damage or wear (cracking, crazing, tears, punctures, etc.). Discard and
replace if necessary.
Note: Refer to the humidier manual for the instructions on how to clean the heated tube.
6.4 Service
The device does not require routine servicing.