Chapter 8 Troubleshooting
67
Question: Why isn’t my detachable battery charging when it is inserted into the Detachable
Battery Module and the ventilator is running on AC power?
Answer: The battery may not charge if the device is too hot or too cold or is operating at an
ambient temperature outside of the specied valid range. Or, the device may not have enough
power to charge the battery if the humidier is in use.
- Make sure the device is not too close to a heat source.
- Ensure the cooling air vents are not blocked.
- Bring the ventilator to ambient room temperature.
- Allow the battery to charge while the device is in Standby or while the airow is on and
humidier is o.
- Use the optional Philips Respironics Detachable Battery Charger to charge your battery.
- If the problem continues, contact an authorized service representative or Philips
Respironics to have the device serviced. Please have the model number and serial number
ready when you call. If you are a patient, please contact your home care service provider.