BiPAP A40 user manual
66
Question: Why is the airow much warmer than usual?
Answer: The air lters may be dirty. Clean or replace the air lters.
- The temperature of the air may vary somewhat based on your room temperature. Make
sure the device is properly ventilated. Keep it away from bedding or curtains that could
block the ow of air around the device.
- Make sure the device is away from direct sunlight and heating equipment.
- If using the humidier with the device, check the humidier settings. Refer to the
humidier instructions to make sure the humidier is working properly.
Question: Why does the mask feel uncomfortable?
Answer: This could be due to improper headgear adjustment or improper mask tting.
- Make sure you are properly tted with the correct size mask.
- If the problem continues, contact your home care service provider to be tted with a
dierent mask.
Question: Why did my prescription change fail when I updated my prescription using the SD
card?
Answer: There are three possible error messages that will appear if the prescription change
fails when using an SD card:
- Prescription Change Failed: Remove the card and have the prescription replaced with a
valid prescription.
- Prescription Failed – Serial Number: Remove the card and have the prescription replaced
with the prescription with the correct serial number.
- Prescription Failed – Version: Remove the card and have the prescription replaced with a
prescription in the correct version.