Chapter 9 Technical Specications
69
Environmental
Operating Storage
Temperature 5° C to 40° C (41° F to 104° F) -20° C to 60° C (-4° F to 140° F)
Relative Humidity 15 to 95% (non-condensing) 15 to 95% (non-condensing)
Atmospheric Pressure 101 kPa to 77 kPa (approximately
0-2286 m/0-7500 ft)
N/A
Physical
Dimensions: 22.23 cm W x 18.42 cm L x 10.80 cm H
(8.75” W x 7.25” L x 4.25” H)
Weight: Approximately 2 kg (4.4 lbs)
Standards Compliance
This device is designed to conform to the following standards:
• IEC 60601-1: Medical Electrical Equipment – Part 1: General Requirements for Basic
Safety and Essential Performance
• IEC 60601-1-2: Medical Electrical Equipment – Part 1-2: General Requirements for Basic
Safety and Essential Performance – Collateral standard: Electromagnetic compatibility -
Requirements and tests
• IEC 60601-1-6: Medical Electrical Equipment – Part 1-6: General Requirements for Safety
and Essential Performance – Collateral standard: Usability
• IEC 60601-1-8: Medical Electrical Equipment – Part 1-8: General Requirements for Basic
Safety and Essential Performance – Collateral standard: General Requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
• IEC 60601-1-11: Medical Electrical Equipment – Part 1-11: General Requirements for Basic
Safety and Essential Performance – Collateral standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare environment
• ISO 10651-6: Lung Ventilators for Medical Use – Particular Requirements for Basic Safety
and Essential Performance, Part 6. Home care ventilatory support devices
• RTCA DO-160F Section 21, Category M; Emission of Radio Frequency Energy
9. Technical Specications