Electrical
AC Voltage Source: 100 to 240 VAC, 50/60 Hz, 1.2 A
DC Power Source: 12 VDC, 5.0 A (External Battery)
24 VDC, 4.2 A (Power Supply)
Type of Protection Against Electric Shock: Class II (To be used with external Class II power
supply only)
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water: Drip Proof, IP22
Mode of Operation: Continuous
SD Card and SD Card Reader
Use only SD cards and SD card readers available from Philips Respironics, including the following:
SanDisk ® Card Reader/Writer - SanDisk ImageMate - REF SDDR-99-A15
Control Accuracy
Parameter Range Accuracy
IPAP 4 – 40 cm H
2
O ± 2.5 cm H
2
O*
EPAP 4 – 25 cm H
2
O ± 2.5 cm H
2
O*
CPAP 4 – 20 cm H
2
O ± 2.5 cm H
2
O*
Breath rate 0 to 40 BPM greater of ± 1 BPM or ±10% of setting
Inspiration time 0.5 to 3 seconds ± (10% of setting + 0.1 second)
Specications listed are based on using a standard patient circuit (Philips Respironics 15 or 22 mm
tubing; Whisper Swivel II).
*Pressure measured at the patient connection port with or without the integrated heated
humidier (no patient ow).